11.1 Introduction
Volumetric analysis is a type of quantitative analysis used to determine the amount of a substance. In quantitative analysis, the unknown amount of a substance is determined with the help of a reaction between a solution containing the unknown amount of the substance and another solution of known concentration. The solution whose concentration is known is called a standard solution.
From the volume of the standard solution required in the reaction and the volume of the solution containing the unknown amount of the substance, the unknown amount can be calculated using the law of chemical equivalence or the molarity equation. The method of determining this is called titration. Therefore, volumetric analysis is also known as titrimetric analysis.
The titration involving an oxidizing agent and a reducing agent is called redox titration.
For any titration, a solution of known concentration (standard solution) is used. To prepare a standard solution, a specific amount of a substance is dissolved in a specific volume of a solvent (usually water). Generally, the amount of a substance is obtained by accurate weighing.
11.2 Chemical balance
A chemical balance is a device or instrument used in a laboratory to accurately weigh a substance. The maximum weight limit that can be weighed in a simple chemical balance is 200g, and the sensitivity of the balance is 0.1 mg. A chemistry student must know about the construction, usage, and precautions to be taken for using a chemical balance.
11.2.1 Construction of chemical balance
The structure of a commonly used chemical balance is shown in Figure 11.1. The balance is kept inside a glass box so that dust and wind cannot damage it. The balance has a thin, but strong and graduated beam. The beam is mounted on a central knife edge. On the other hand, the central knife edge rests on an agate plate at the top of the balance column. Two pans are suspended from the two terminal knife edges, which are at equal distances from the central knife edge, using stirrups.
A long pointer is connected to the midpoint of the beam. The pointer can move freely back and forth over a pointer scale at the base of the beam.
To use the balance accurately, the central knife edge and the two terminal knife edges must be in the same plane. An arrest knob is used for the arresting arrangement of the balance. Lifting the beam up with the help of the knob allows the long pointer to move. This arrangement is activated to take the actual reading of the weight. At other times, the knob is kept in its normal position, keeping the balance in an arrested state.
Leveling screws are attached to the base of the balance. The balance is leveled using these two screws. Adjusting screws are located at the two ends of the graduated beam. By using these two screws, the long pointer can be made to swing equally on the pointer scale. That is, the arrangement to ensure that the pointer crosses an equal number of marks on both sides of the zero mark of the pointer scale is done using the adjusting screws.
To increase the accuracy of weighing of a balance beyond 0.01 g, a rider is used. The rider is used with the help of a rider-hook of the rider carriage. The rider is a thin, slender aluminum wire with a bend. The weight of this wire is 0.01 or 0.005 g. The arrangement of marks on the beam and the weight of the rider depend on the type of chemical balance. In a Sartorius balance, a 0.01 g rider is used, and the zero mark is given at the midpoint of the graduated beam. There are 50 equal divisions on both sides of the zero mark. This provides the rider scale. For each mark on the right side of the beam, a weight of 0.0002 g should be added. On the other hand, for each mark on the left side, 0.0002 g is subtracted. Such weight should be calculated depending on the position of the rider.
11.2.2 Weight box
A wooden box containing weights of various denominations is used for weighing. This is called a weight box (Figure 11.2). The weights in grams are available as 100, 50, 20, 10, 5, 2, and 1g. The weights representing fractional values of a gram are called fractional weights. A weight box contains weights of 500, 200, 100, 50, 20, and 10 mg. Usually, a 10 mg rider is also present in the box. A forcep is provided in the box to use the weights and the rider.
By using the weights, weighing can be done up to an accuracy of 0.01 g. For higher accuracy, a rider is used.
11.2.3 Steps involved in weighing
- Bring the balance to a leveled position using the leveling screws. In the leveled state, the long pointer rests at the zero mark of the pointer scale. The balance should be in the arrested state during this time.
- Now, lift the beam up using the arrest knob and observe the swing of the pointer. If necessary, use the adjusting screws to ensure the pointer swings equally on both sides of the zero mark of the pointer scale. To make this adjustment, i.e., to use the adjusting screws, the balance must always be brought to the arrested state.
- A weighing bottle or watch glass is generally used to weigh a substance. Place a clean weighing bottle or watch glass in the center of the left pan. Do this only while keeping the balance in the arrested state.
- Now, place appropriate weights on the right pan and finally complete the weighing using the rider. When the weights on both pans are equal, the pointer will swing equally on both sides of the zero mark. Remember that the balance must be kept in the arrested state every time a weight is placed on the pan or the position of the rider is changed.
- Record the value of the weights on the right pan and the value of the weight to be added or subtracted due to the rider's position.
- Now, using appropriate weights on the right pan and placing the rider in the appropriate position, add the required amount of substance (by weight) to the watch glass or weighing bottle on the left pan.
- Now, take the watch glass or weighing bottle with the substance out of the balance from the left pan and transfer the substance to the appropriate place.
11.2.4 Precautions for using the balance
- Keep the balance in an appropriate place and do not keep moving it from there.
- Do not weigh a substance when it is hot. Bring the temperature of the substance to normal temperature before weighing.
- Before using the balance, check if the balance is functioning properly. Ensure that the pointer of the balance can swing freely.
- The balance must be in the arrested state during various steps such as adding weights, removing weights, changing the position of the rider, placing a watch glass on the pan, adding substance to the watch glass, etc.
- The arrest knob should be used smoothly to lift the beam.
- A forcep must always be used to handle the weights and the rider. Never place weights or the rider on the table or the floor of the balance.
- Any substance or weight should always be placed in the center of the pan.
- If any substance falls on the floor inside the balance or on the pan, remove it immediately and clean the pan or the floor.
- The values of the weights obtained during weighing and the values obtained from the rider's position must always be properly recorded.
11.3 Concentration of solution
If the concentration of a solution is known, it is called a standard solution. Thus, the amounts of solute and solvent in a standard solution are known. To prepare a standard solution, a specific weight of solute is usually dissolved in a specific volume of solution or solvent.
For pure and stable compounds like oxalic acid, sodium carbonate, Mohr's salt, etc., a specific weight can be accurately measured and dissolved in a specific volume of solution to prepare a standard solution. Such substances are called primary standards.
On the other hand, substances like HCl, H2SO4, NaOH, etc., are not found in a pure state and are hygroscopic. Therefore, such substances cannot be accurately weighed. These substances are called secondary standards. For such substances, an approximate specific amount is weighed and a solution is prepared. The exact concentration of this solution is determined by titration using a standard solution, and consequently, such a solution can be used as a standard solution.
Although there are various units to express the concentration of a solution, usually in the case of neutralization reactions or redox reactions, the concentration is expressed by normality and molarity.
The number of equivalents of solute present in one liter of a solution is called the normality of the solution. Therefore,
Normality =
The number of moles of solute present in one liter of a solution is called the molarity of the solution. Therefore,
Molarity =
11.4 Equivalent mass of oxidizing and reducing agent
In a redox reaction, a reaction occurs between an oxidizing agent and a reducing agent. In such a reaction, the oxidizing agent itself gets reduced, and the reducing agent itself gets oxidized. Electrons are added in reduction, and electrons are released in oxidation. The mass of a substance (oxidizing or reducing agent) participating in a redox reaction that accepts or releases one mole of electrons is called the equivalent mass of the substance. Mathematically, this amount is equal to the molar mass of the substance divided by the number of electrons accepted or released by one molecule of the substance in the reaction. Therefore,
Equivalent mass of oxidizing agent =
Equivalent mass of reducing agent =
The number of electrons can be known from the reduction half-reaction of the oxidizing agent or the oxidation half-reaction of the reducing agent in the redox reaction.
For example, the reduction half-reaction of KMnO4 as an oxidizing agent in the presence of dilute H2SO4 is:
MnO4- + 8H+ + 5e- → Mn2+ + 4H2O
Therefore, Equivalent mass of KMnO4 (in acidic medium) =
Similarly, the oxidation half-reaction of the reducing agent oxalic acid (the formula of pure oxalic acid is H2C2O4.2H2O) is:
C2O42- → 2CO2 + 2e-
Therefore, Equivalent mass of oxalic acid =
The half-reaction of the reducing agent ferrous sulphate (FeSO4.7H2O) or ferrous ammonium sulphate (or Mohr's salt) [FeSO4.(NH4)2SO4.6H2O] is:
Fe2+ → Fe3+ + e-
Therefore, Equivalent mass of ferrous sulphate =
And, Equivalent mass of Mohr's salt =
11.5 Law of chemical equivalence and molarity equation
By using the law of chemical equivalence or the molarity equation, the amount of oxidizing or reducing agent can be calculated with the help of a redox titration.
11.5.1 Law of chemical equivalence
According to this law, at the end point of a titration, the number of equivalents of the substance to be titrated (called the titre) is equal to the number of equivalents of the titrant used. Suppose, for the end point of a redox titration, V1 mL of an oxidizing agent of normality N1 reacts with V2 mL of a reducing agent of normality N2.
We know that, 1000 mL 1N solution of an oxidizing agent = 1 equivalent of the oxidizing agent
∴ V1 mL N1 normal solution of the oxidizing agent =
Similarly, V2 mL N2 normal solution of the reducing agent =
At the end point of the titration,
V1N1 = V2N2 (11.5)
Equation 11.5 is the fundamental equation of volumetric analysis. This equation can be called the normality equation.
11.5.2 Molarity equation
Nowadays, the concept of molarity is being increasingly used. Therefore, calculations based on the mole concept have become more necessary. By using the molarity equation, calculations for titrations can be done using the mole concept. The molarity equation is obtained with the help of the stoichiometric coefficients of the balanced equation for a redox reaction. Suppose, the following equation is the balanced equation for a redox reaction.
n1A + n2B → n3C + n4D
Suppose, in the reaction, V1 mL of oxidizing agent A of molarity M1 reacts with V2 mL of reducing agent B of molarity M2.
Now, V1 mL of oxidizing agent A of molarity M1 contains
Similarly, V2 mL of reducing agent B of molarity M2 contains
From the balanced equation, it is known that n1 moles of oxidizing agent A completely react with n2 moles of reducing agent B.
∴
or
Equation 11.6 is the molarity equation for redox titration.
The balanced equation for the redox reaction involving the oxidation of ferrous sulphate by KMnO4 in an acidic medium is:
2KMnO4 + 10FeSO4 + 8H2SO4 → K2SO4 + 2MnSO4 + 5Fe2(SO4)3 + 8H2O
Therefore, according to the molarity equation, n1 = 2 (coefficient of oxidizing agent) and n2 = 10 (coefficient of reducing agent). Thus, the molarity equation for the redox reaction will be:
or
This molarity equation will also be applicable for the redox reaction between ferrous ammonium sulphate (Mohr's salt) and KMnO4 in an acidic medium.
Similarly, the balanced equation for the redox reaction involving the oxidation of oxalic acid by KMnO4 in an acidic medium is:
2KMnO4 + 5H2C2O4 + 3H2SO4 → K2SO4 + 2MnSO4 + 8H2O + 10CO2
Here, n1 = 2 and n2 = 5. Therefore, the molarity equation for the redox reaction will be:
Thus, if the values of three quantities in the molarity equation are known, the value of the fourth quantity can be calculated.
11.6 End point of a titration and indicator
The point at which the reaction between the two solutions used in a titration is complete is called the end point of the titration. The substance that can indicate the end point of a titration is called an indicator. Indicators show a change in color at the end point.
The indicators used in redox titrations are usually certain organic compounds, whose oxidized state and reduced state have different colors. For example, in the titration between Fe2+ and acidic K2Cr2O7 solution, diphenylamine is used as an indicator, and it shows a color change from green to purple.
In the titration of KMnO4 in an acidic medium, a separate indicator is not required. In such titrations, the pink-colored KMnO4 (MnO4- ion) acts as a self-indicator. By reacting with the reducing agent in an acidic medium, the MnO4- ion of KMnO4 is reduced to colorless Mn2+. In a titration, KMnO4 solution is added dropwise to the reducing agent solution. The pink color of each drop becomes colorless due to the reaction with the reducing agent. When the added KMnO4 solution completely reacts with the reducing agent (the end point of the titration), one extra drop of KMnO4 solution turns the entire solution pink.
11.7 Apparatus used in volumetric analysis
In a titration, a burette, pipette, and a titration flask (usually a conical flask) are used.
11.7.1 Burette
A burette is made of a cylindrical glass tube of uniform radius. Burettes of different volumes are available. In the laboratory, a 50 mL burette is commonly used. The lower part of the burette is narrow, and the very bottom part has a jet. A stop cock is attached to the lower part. In some burettes, a glass jet is attached to the lower part using a rubber tube. A clip is attached to the rubber tube. When the stop cock or clip is closed, the solution cannot flow out of the burette.
The main tube of the burette has 50 equal divisions within the 50 mL volume. Marks are drawn on the burette, writing the numbers sequentially from 0 (zero) at the top to 50 at the bottom. The volume of each of these divisions is 1 mL. Each division is further divided into 10 equal smaller divisions with marks. The volume of each smaller division is 0.1 mL.
Handling of burette:
Take a clean burette and wash it first with water. Now, fill 3-5 mL of the solution to be taken in the burette, hold the burette horizontally, and slowly rotate it to rinse the inside with the solution. Open the stop cock and let this solution flow out. Now, completely fill the burette with the solution to be used in the experiment and fix it upright on a stand using a clamp. Ensure that there are no empty spaces due to air bubbles in the solution inside the burette. If there are air bubbles, open the stop cock completely and let the solution flow out forcefully. Adjust the lower meniscus of the solution, which should be free of empty spaces due to air bubbles, to the zero mark.
Reading of burette:
If the solution taken in the burette is colorless, align the lower meniscus of the solution with the zero mark of the burette. For colored solutions, the lower meniscus is not clearly visible. Therefore, for such solutions, the upper meniscus is aligned with the zero mark. To take an accurate burette reading after the experiment, an anti-parallax arrangement should be adopted. For this, take a thick piece of paper or card (12 cm x 5 cm), fold it in half, and make two cuts at a distance of 5 cm as shown in the figure. Now, open the fold of the card and slip the paper onto the burette using the two cuts. This arrangement can eliminate errors in the burette reading due to parallax. When taking the burette reading, the eye should be kept at the level of the meniscus of the solution, so that an accurate reading can be taken with the help of the thick paper arrangement.
11.7.2 Pipette
A pipette is a long glass tube with a bulb in the middle and a jet at the very bottom. There is a mark on the tube part above the bulb. Up to this mark, the volume of solution the pipette holds can be transferred using the pipette. The value of this volume is written on the bulb of the pipette. Pipettes of various volumes are available, such as 10 mL, 20 mL, 25 mL, 50 mL, etc.
Handling of pipette:
Take a cleaned pipette and wash it again with water. Now, take a little bit of the solution to be used in the pipette and rinse it. To do this, hold the pipette horizontally and rotate it so that the solution wets all parts of the inside. Discard the used solution. Now, dip the lower part of the pipette into the solution and suck the solution up through the upper mouth, letting it rise slightly above the mark above the bulb. Immediately close the upper mouth tightly with your index finger. Now, hold the pipette upright, slightly loosen the pressure of your finger, and let the solution come out drop by drop from the pipette. The moment the lower meniscus of the solution (upper meniscus for colored solutions) is on the same plane as the mark on the pipette, immediately press your finger down again to stop the solution from coming out. Then, insert the pointed end of the pipette into a clean titration flask and release the pressure of your finger. After all the solution has been transferred, touch the tip of the pipette to the inner wall of the flask once or twice. This is done to transfer the last drop (of the specific volume) from the pipette. Even after this, never try to transfer the solution remaining in the tip of the pipette by blowing or any other means.
11.7.3 Titration flask
Usually, a 250 mL conical flask is used as a titration flask. Before using it, it should be washed clean with water. Finally, it should be rinsed with distilled water. (A titration flask should never be rinsed with the solution used in the titration.)
11.8 Description of titration
In a redox titration using KMnO4 solution, the KMnO4 solution is taken in the burette and the reducing agent solution is taken in the titration flask. The burette, pipette, and titration flask (conical flask) must first be cleaned and washed thoroughly with water.
Then, the burette is rinsed with the KMnO4 solution and the pipette is rinsed with the reducing agent solution. The rinsed burette is filled with the KMnO4 solution and fixed upright on a stand using a clamp. The stop cock is opened as needed to align the upper meniscus of the KMnO4 solution with the zero mark. Meanwhile, a specific volume of the reducing agent solution is filled into the titration flask using the pipette. As required, dilute H2SO4 is added to the reducing agent solution in the flask. Now, the flask is placed on a piece of white paper under the burette. Then, the stop cock is opened, and the KMnO4 solution from the burette is allowed to fall drop by drop into the flask. The stop cock should be operated with the left hand, and the flask should be held and swirled with the right hand so that the added KMnO4 and the reducing agent solution mix well. At a certain time, the end point of the titration will be reached, and one extra drop of KMnO4 solution will permanently turn the entire solution a light pink color. Immediately, the stop cock is closed, and the burette reading is taken. This burette reading is the final burette reading. The difference between the initial burette reading (usually zero) and the final burette reading is the volume of KMnO4 solution required, which completely reacts with the specific volume of the reducing agent taken in the titration flask. Now, discard the solution from the flask and wash the flask thoroughly. Finally, rinse the flask with distilled water and repeat the whole process again, and find out the volume of KMnO4 solution again from the burette reading. The process should be repeated until the volume of KMnO4 solution obtained from two experiment readings is identical (a difference of 0.05 mL is acceptable).
Some points to note during titration:
- Wash the burette, pipette, and titration flask thoroughly with water.
- Rinse the burette with the solution to be taken in the burette.
- Rinse the pipette with the solution to be transferred using the pipette.
- Rinse the titration flask with distilled water.
- Always ensure there is no empty space due to the presence of air in the tip of the burette. Open the stop cock completely and let the solution come out forcefully to completely eliminate such empty space.
- To eliminate errors in the burette reading due to parallax, adopt the anti-parallax arrangement.
- Use the lower meniscus for colorless solutions and the upper meniscus for colored solutions to align with the marks on the burette or pipette.
- Keep your eye at the level of the meniscus of the solution while taking the burette reading.
- Never transfer the last drop remaining in the jet of the pipette by blowing or any other means.
- Never use an excess amount of indicator.
- Keep swirling the titration flask well during the titration so that the mixture inside mixes well.
- The titrant solution should not stick to the inner wall of the flask. If this happens, mix such drops with the main mixture using water.
- Let the drops fall slowly from the burette just before reaching the end point.
- Do not keep touching the main body of the burette or pipette with your hand during the experiment, because body heat expands the glass.
- If KMnO4 solution is used in the burette, do not use a burette with a jet attached to a rubber tube, because KMnO4 reacts with rubber.
Experiment 11.1: Preparation of 250 mL of standard (N/10) solution of oxalic acid.
Theory: Oxalic acid is a primary standard substance. Its molecular formula in the pure state is H2C2O4.2H2O. Therefore, its molecular mass is 126. The oxidation half-reaction of oxalic acid is:
C2O42- → 2CO2 + 2e-
Therefore, Equivalent mass of oxalic acid =
∴ Amount of oxalic acid required for 250 mL N/10 solution =
Apparatus: Watch glass, beaker, funnel, glass rod, 250 mL volumetric flask, wash bottle, weight box, and chemical balance.
Chemicals: Pure oxalic acid and distilled water.
Procedure:
- Take a clean and dry watch glass.
- Weigh the watch glass accurately and record the weight value in the notebook (W1 g).
- Take a specific weight of pure oxalic acid on the watch glass and weigh it. The difference between the weight of the watch glass and the weight of the watch glass with oxalic acid should be between 1.5 - 1.6 g. Record this weight value as well (W2 g).
- Carefully transfer the oxalic acid from the watch glass to a 250 mL beaker. Wash the watch glass with distilled water from the wash bottle as shown in Figure 11.10, so that the oxalic acid particles sticking to the glass go into the beaker.
- Stir with a glass rod and dissolve the oxalic acid in the water in the beaker.
- Transfer the oxalic acid solution from the beaker to the volumetric flask using a funnel (Figure 11.11).
- Wash the inside of the beaker thoroughly with distilled water and transfer the water to the volumetric flask (Figure 11.12).
- Wash the inside of the funnel with distilled water and let the washings go into the volumetric flask.
- Now carefully add distilled water to the volumetric flask so that the water fills almost up to the mark on the flask. To bring it to the exact volume, add distilled water drop by drop using a pipette or dropper so that the lower meniscus of the solution touches the mark on the flask.
- Close the volumetric flask with an appropriate stopper and carefully shake the flask so that the solution inside becomes homogeneous.
- Calculate the exact normality of the solution from the actual exact weight of the oxalic acid and stick a label with the normality value on the flask.
Observations and calculations:
- Weight of watch glass = W1 g
- Weight of watch glass with oxalic acid = W2 g
- Volume of prepared solution = 250 mL
- Weight of oxalic acid in the solution = (W2 - W1) g
Normality of oxalic acid solution =
Experiment 11.2: Preparation of 250 mL of standard (M/10) solution of oxalic acid.
Theory: Pure oxalic acid is H2C2O4.2H2O. Its molecular mass is 126.
∴ Amount of oxalic acid required for 250 mL standard M/10 solution =
Apparatus: Same as Experiment 11.1.
Chemicals: Same as Experiment 11.1.
Procedure: Almost the same as Experiment 11.1. Only in step (iii), the difference in weight between the watch glass and the watch glass with oxalic acid should be between 3.1 and 3.2 g.
Observations and calculations: Same as Experiment 11.1. Only the calculation in the last line will be:
Molarity of oxalic acid =
Experiment 11.3: Preparation of 250 mL of standard (M/20) solution of Mohr's salt.
Theory: Mohr's salt is a primary standard substance. Its molecular formula is FeSO4.(NH4)2SO4.6H2O and its molecular mass is 392.
∴ Amount of Mohr's salt required for 250 mL M/20 solution =
Apparatus: Same as Experiment 11.1.
Chemicals: Pure Mohr's salt, pure concentrated H2SO4, and distilled water.
Procedure:
- Accurately weigh about 4.9 g of Mohr's salt as described in Experiment 11.1 (the difference in weight between the watch glass and the watch glass with Mohr's salt should be between 4.8 g and 5 g).
- Carefully transfer the weighed Mohr's salt to a 250 mL beaker as described in Experiment 11.1 and add about 5 mL of concentrated H2SO4 to it (H2SO4 is added to prevent the hydrolysis of ferrous sulphate).
- Now complete the experiment as described in Experiment 11.1.
Observations and calculations: Same as Experiment 11.1 and the relation required for the calculation in the last line is:
Molarity of Mohr's salt solution =
Experiment 11.4: Preparation of 250 mL of standard (N/50) solution of Mohr's salt.
Theory: The molecular mass of Mohr's salt [FeSO4.(NH4)2SO4.6H2O] is 392. The oxidation half-reaction of this reducing agent is:
Fe2+ → Fe3+ + e-
Therefore, the equivalent mass of Mohr's salt is equal to its molecular mass.
∴ Amount of Mohr's salt required for 250 mL N/50 solution =
Apparatus: Same as Experiment 11.3.
Chemicals: Same as Experiment 11.3.
Procedure: Perform the experiment as described in Experiment 11.3. The difference in weight between the watch glass and the watch glass with Mohr's salt should be between 1.9 and 2 g.
Observations and calculations: Same as Experiment 11.3 and for the calculation in the last line we get,
Normality of Mohr's salt solution =
Experiment 11.5: Preparation of 250 mL of a standard (M/50) solution of oxalic acid and with the help of the standard solution determination of molarity and g L-1 of a given solution of KMnO4.
Theory: Oxalic acid reacts with KMnO4 in the presence of dilute H2SO4. The balanced equation obtained for this reaction is:
2KMnO4 + 5H2C2O4 + 3H2SO4 → K2SO4 + 2MnSO4 + 8H2O + 10CO2
The ionic equation obtained for the reaction is:
2MnO4- + 5C2O42- + 16H+ → 2Mn2+ + 10CO2 + 8H2O
In the reaction, oxalic acid acts as a reducing agent and KMnO4 acts as an oxidizing agent. In the titration between KMnO4 and oxalic acid, KMnO4 acts as a self-indicator and the creation of a permanent light pink color indicates the end point.
Apparatus: Chemical balance, weight box, watch glass, 250 mL beaker, glass rod, 250 mL volumetric flask, funnel, titration flask, wash bottle, wire gauze, burette.
Chemicals: Pure oxalic acid (H2C2O4.2H2O), 4N H2SO4, and the KMnO4 solution to be tested (the strength of the solution should be approximately M/125 or 1.264 g L-1).
Procedure:
- Accurately weigh about 0.63 g of pure oxalic acid and prepare 250 mL of standard M/50 oxalic acid solution as described in Experiment 11.1.
- Take a clean burette with a stop cock. Rinse the burette with the KMnO4 solution to be tested, fill it with KMnO4, and fix it upright on a stand using a clamp. Open the stop cock of the burette as needed to align the upper meniscus of the KMnO4 solution with the zero mark of the burette.
- Transfer 25 mL of oxalic acid to a clean titration flask using a pipette rinsed with oxalic acid. Now add about 20 mL of 4N H2SO4 to the oxalic acid solution and heat the mixture solution to about 60°-70°C using a wire gauze.
- Now bring the flask under the burette, add two or three drops of KMnO4 solution from the burette into it, and shake well. After the pink color of KMnO4 disappears, keep adding KMnO4 solution drop by drop and shake the mixture every time. If you keep adding drop by drop like this, at a certain time one drop of KMnO4 solution will permanently turn the whole solution a light pink color. This is the end point of the titration. Take the final burette reading (y) from the position of the upper meniscus of the KMnO4 solution in the burette.
- The whole experiment should be repeated two or three times to obtain two identical burette readings.
Observations:
- Molarity of oxalic acid solution =
(W2 - W1) × 1000 / 250 × 126 = M2 - Volume of oxalic acid solution = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
The molarity equation obtained from the balanced equation of the redox reaction between KMnO4 and oxalic acid in an acidic medium is:
∴ M1 (Molarity of KMnO4 solution) =
=
∴ Strength of KMnO4 solution in g L-1 = M1 × 158 g L-1
Experiment 11.6: Preparation of 250 mL of a standard (N/10) solution of oxalic acid and with the help of the standard solution determination of normality and g L-1 of a given solution of KMnO4.
Theory: Same as Experiment 11.5.
Chemicals: Same as Experiment 11.5 (The strength of the given KMnO4 solution should be approximately N/10, i.e., it should contain about
Procedure: Prepare 250 mL of standard (N/10) oxalic acid solution as described in Experiment 11.1.
After that, follow the procedure described in Experiment 11.5.
Observations:
- Normality of oxalic acid solution =
(W2 - W1) × 1000 / 250 × 63 = N2 - Volume of oxalic acid solution = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
From the law of chemical equivalence,
V1N1 = V2N2 (See Equation 11.5)
∴ N1 (Normality of KMnO4 solution) =
=
Therefore, g L-1 of KMnO4 solution = Normality × Equivalent mass = N1 × 31.6 g L-1
(∴ Equivalent mass of KMnO4 = 31.6)
Experiment 11.7: Preparation of 250 mL of a standard (M/20) solution of Ferrous ammonium sulphate (Mohr's salt) and with the help of the standard solution determination of the strength of a given KMnO4 solution in terms of molarity and g L-1.
Theory: In an acidic medium, the MnO4- ion easily oxidizes the Fe2+ ion to form Fe3+. The ionic equation obtained for this is:
MnO4- + 5Fe2+ + 8H+ → Mn2+ + 5Fe3+ + 4H2O
The molecular equation for the reaction is:
2KMnO4 + 10FeSO4.(NH4)2SO4.6H2O + 8H2SO4 → K2SO4 + 2MnSO4 + 5Fe2(SO4)3 + 10(NH4)2SO4 + 68H2O
In the titration between KMnO4 and Mohr's salt, KMnO4 acts as a self-indicator and the creation of a permanent light pink color indicates the end point.
Apparatus: Chemical balance, weight box, watch glass, 250 mL beaker, glass rod, 250 mL volumetric flask, funnel, titration flask, wash bottle, burette.
Chemicals: Pure Mohr's salt, concentrated H2SO4, 4N H2SO4, and the KMnO4 solution to be tested (The strength of the KMnO4 solution should be approximately M/100 or 1.58 g L-1).
Procedure:
- Prepare 250 mL of standard (M/20) Mohr's salt solution as described in Experiment 11.3.
- Fill the burette with KMnO4 solution as given in Experiment 11.5.
- Transfer 25 mL of Mohr's salt solution to a clean titration flask using a pipette rinsed with Mohr's salt solution. Now add about 20 mL of 4N H2SO4 to the solution in the flask.
- Keep adding KMnO4 solution drop by drop from the burette to the flask and shake the mixture every time. If you keep adding drop by drop like this, at a certain time one drop of KMnO4 solution will permanently turn the whole solution a light pink color. This is the end point of the titration. Take the final burette reading (y) from the position of the upper meniscus of the KMnO4 solution in the burette.
- The whole experiment should be repeated two or three times to obtain two identical burette readings.
Observations:
- Molarity of Mohr's salt solution =
(W2 - W1) × 1000 / 250 × 392 = M2 - Volume of Mohr's salt solution = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
The molarity equation obtained from the balanced equation of the redox reaction between KMnO4 and Mohr's salt in an acidic medium is:
∴ M1 (Molarity of KMnO4 solution) =
=
∴ Strength of KMnO4 solution in g L-1 = M1 × 158 g L-1
Experiment 11.8: Preparation of a solution of Mohr's salt having the concentration of 17 g L-1 and determination of the strength of a given KMnO4 solution in terms of molarity and g L-1 with the help of the prepared solution of Mohr's salt.
Theory: Same as Experiment 11.7.
Apparatus: Same as Experiment 11.7.
Chemicals: Same as Experiment 11.7 (The strength of the given KMnO4 solution should be approximately M/100 or 1.58 g L-1).
Procedure: Accurately weigh
After that, follow the procedure described in Experiment 11.7.
Observations:
- Molarity of Mohr's salt solution =
17 / 392 = 0.043 M = M2 - Volume of Mohr's salt solution = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
The molarity equation obtained for the redox reaction between KMnO4 and Mohr's salt in an acidic medium is:
∴ M1 (Molarity of KMnO4 solution) =
∴ M1 (g L-1 of KMnO4 solution) = M1 × 158 =
Experiment 11.9: Preparation of 250 mL of a standard (M/20) ferrous ammonium sulphate (Mohr's salt) solution and with the help of the standard solution, determination of percentage purity of an impure sample of KMnO4 of 2 g, dissolved in 1 L of a solution.
Theory: Same as Experiment 11.5.
Chemicals: Pure Mohr's salt, concentrated H2SO4, 4N H2SO4, and the given KMnO4 solution (2 g dissolved in 1 L solution).
Procedure: Prepare 250 mL of standard (M/20) Mohr's salt solution as described in Experiment 11.3.
After that, follow the procedure described in Experiment 11.7.
Observations:
- Molarity of Mohr's salt =
(W2 - W1) × 1000 / 250 × 392 = M2 - Volume of Mohr's salt = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
According to the molarity equation obtained for the redox reaction between KMnO4 and Mohr's salt in an acidic medium,
∴ M1 (Molarity of KMnO4 solution) =
∴ g L-1 of KMnO4 solution = M1 × 158 g L-1
Therefore, Percentage purity =
Experiment 11.10: Preparation of 250 mL of a standard (M/40) oxalic acid solution and with the help of the standard solution, determination of percentage purity of a sample of impure KMnO4 by dissolving 2 g of the sample in 1L solution.
Theory: Same as Experiment 11.5.
Apparatus: Same as Experiment 11.5.
Chemicals: Pure oxalic acid (H2C2O4.2H2O), 4N H2SO4, and the given KMnO4 solution (2 g dissolved in 1 L solution).
Procedure: Accurately weigh about 0.7875 g (=
After that, follow the procedure described in Experiment 11.5.
Observations:
- Molarity of oxalic acid =
(W2 - W1) × 1000 / 250 × 126 = M2 - Volume of oxalic acid = 25 mL = V2
Volume of KMnO4 solution (V1) from burette reading:
| Serial No. | Burette reading | Volume of KMnO4 solution used = y - x = V1 (mL) | |
|---|---|---|---|
| Initial reading (x) | Final reading (y) | ||
| 1 | |||
| 2 | |||
| 3 | |||
| 4 | |||
Calculations:
According to the molarity equation obtained for the redox reaction between KMnO4 and oxalic acid in an acidic medium,
∴ M1 (Molarity of KMnO4 solution) =
∴ g L-1 of KMnO4 solution = M1 × 158
Therefore, Percentage purity =
Some important points to know for titrations where KMnO4 is used as an oxidizing agent in an acidic medium:
- (i) The equivalent mass of KMnO4 is 1/5th of its molecular mass.
- (ii) KMnO4 is not a primary standard. Because KMnO4 is not found in a completely pure state.
- (iii) KMnO4 reacts with rubber. Therefore, a burette with a rubber stop cock should not be used in KMnO4 titrations. In KMnO4 titrations, the KMnO4 solution is taken in the burette.
- (iv) Before starting the titration, approximately an equal volume of 4N H2SO4 is mixed with the reducing agent solution taken in the titration flask.
- (v) If the reducing agent solution is oxalic acid, the oxalic acid solution containing 4N H2SO4 is first heated to a temperature of 60°-70°C. Only then should the titration be started. The solution is heated in this way to increase the rate of the reaction.
- (vi) In the titration of KMnO4 and oxalic acid in the presence of dilute H2SO4, the MnSO4 produced as a result of the reaction acts as an auto-catalyst.
- (vii) Ferrous sulphate solution should never be heated while titrating with KMnO4 solution.
- (viii) KMnO4 acts as a self-indicator.
- (ix) The upper meniscus of the KMnO4 solution should be used for burette readings.
Viva Questions and Answers
Question 1: What is a standard solution?
Answer: A solution whose concentration is known is called a standard solution.
Question 2: What is meant by volumetric analysis?
Answer: The quantitative determination of the amount of solute in a solution with the help of a reaction between a specific volume of that solution and a specific volume of another standard solution is called volumetric analysis.
Question 3: What is a primary standard?
Answer: Substances that can be directly weighed accurately and used for the preparation of standard solutions are called primary standard substances. Primary standard substances are found in a stable and pure state.
Question 4: Why is KMnO4 not a primary standard?
Answer: KMnO4 is not found in a completely pure state. Also, the aqueous solution of KMnO4 is not stable.
Question 5: What is the relationship between the molecular mass and equivalent mass of oxalic acid (as a reducing agent)?
Answer: For oxalic acid,
Normality =
Therefore, Normality = 2 x Molarity
Question 6: What is the end point of a titration?
Answer: The point at which the reaction between the two solutions used in a titration is complete is called the end point of the titration.
Question 7: What is an indicator?
Answer: An indicator is a chemical substance that can indicate the end point of a titration by a change in color.
Question 8: What is a titrant?
Answer: The solution used in the burette in a titration is called a titrant.
Question 9: Why is no external indicator used in redox titrations where KMnO4 is used as a titrant?
Answer: Because in such titrations KMnO4 acts as a self-indicator. The end point of such a titration is obtained by the creation of a permanent light pink color.
Question 10: Why is a titration flask not rinsed with the solution to be taken in the flask?
Answer: After rinsing, some part of the solution will remain attached to the inside of the flask. Therefore, the amount of solution in the flask will become slightly more than the amount transferred by the pipette.
Question 11: Why is a burette rinsed with the titrant?
Answer: It is rinsed in this way so that the concentration of the titrant does not change.
Question 12: Why is it necessary to remove air bubbles from the jet of the burette containing the titrant before starting a titration?
Answer: Because if the air bubbles disappear during the titration, an accurate burette reading will not be obtained.
Question 13: Why is the upper meniscus considered for the burette reading of KMnO4?
Answer: The lower meniscus of KMnO4 is not clearly visible.
Question 14: In the titration of KMnO4 and oxalic acid, why is the acidified oxalic acid solution heated to a temperature of 60°-70°C before starting the titration?
Answer: Because the rate of the reaction is very slow at normal temperature, such temperature is required.
Question 15: In the titration of KMnO4 and oxalic acid, why does the reaction become very fast after the reaction of the initial two or three drops of KMnO4 solution is complete?
Answer: The MnSO4 produced as a result of the initial reaction acts as an auto-catalyst and increases the rate of the reaction.
Question 16: Why is dilute H2SO4 used instead of HNO3 or HCl to use KMnO4 as an oxidizing agent in an acidic medium?
Answer: Because HNO3 is itself an oxidizing agent. On the other hand, HCl has reducing power, due to which it can react with KMnO4.
Question 17: Why should H2SO4 be added along with the solution in the preparation of standard solution of Mohr's salt?
Answer: H2SO4 is added to stop the hydrolysis of ferrous sulphate in Mohr's salt.
Question 18: In the case of oxalic acid, the equivalent mass is half of the molecular mass. On the other hand, in the case of Mohr's salt, the equivalent mass is equal to the molecular mass. Why?
Answer: The half-reaction of oxalic acid as a reducing agent is:
C2O42- → 2CO2 + 2e-
That is, each molecule releases two electrons, for which the equivalent mass is half of the molecular mass. On the other hand, in the case of Mohr's salt, the half-reaction is,
Fe2+ → Fe3+ + e-
This time, each molecule releases one electron, hence the equivalent mass is equal to the molecular mass.
Question 19: Why cannot a burette with a rubber pinch cock be used in the titration of KMnO4?
Answer: KMnO4 can react with rubber, which creates the possibility of a change in the concentration of the KMnO4 solution.
Question 20: 25 mL of 0.1N Mohr's salt solution reacts with 20 mL of a KMnO4 solution in a titration. Calculate the concentration of the KMnO4 solution in normality and g L-1.
Answer: V1N1 = V2N2
∴ N1 =
Therefore, Normality of KMnO4 solution = 0.125N
Now, g L-1 = Normality × Equivalent mass
= 0.125 × 31.6 = 3.95 g L-1